Engineering · 12 min
Compressed Air Quality Verification — ISO 8573-4, -5, and -7
Measurement Methods for Solid Particles, Oil Content, and Viable Microbiological Contamination in Compressed Air Systems
ISO 8573-1:2010 establishes the purity class system for compressed air quality. Verification that installed filtration, drying, and treatment equipment achieves the specified purity class requires measurement methods defined in the ISO 8573 Part 2 through Part 7 series. This article covers the three measurement disciplines most relevant to compressed air system commissioning and periodic audit: solid particle content measurement (ISO 8573-4:2019), oil aerosol and vapour content measurement (ISO 8573-2:2018 and ISO 8573-5:2001), and viable microbiological contamination measurement (ISO 8573-7:2003). Each method specifies sampling conditions, equipment requirements, and calculation procedures that must be followed to produce valid, comparable results. The related article covering ISO 8573-1 purity classes, dryer selection, and system design is available at the compressed-air-purity and compressed-air-dryer-selection engineering references.
±5% of main airstream velocity (ISO 8573-4)
Isokinetic Velocity Tolerance
≤0.01 mg/m³ total oil (ISO 8573-1)
Oil Aerosol Class 1 Limit
Sufficient for ≤0.01 mg/m³ Class 1 verification
Oil Aerosol Detection (GC-FID)
≤0.1 CFU/m³ (ISO 8573-7)
Microbiological Class W1
≥1,000 L at 12.5 L/min
Microbiological Minimum Sample
≥1,000 L gravimetric (ISO 8573-4)
Solid Particle Min Sample
01 /
Isokinetic Sampling — The Fundamental Requirement
All compressed air quality measurements require isokinetic sampling: the velocity of gas entering the sample probe must equal the velocity of the main airstream at the sample point. When probe velocity exceeds airstream velocity (super-isokinetic), larger particles are deflected around the probe inlet, under-representing heavy contamination. When probe velocity is below airstream velocity (sub-isokinetic), larger particles are over-collected. ISO 8573-4 requires isokinetic sampling with a velocity match within ±5% of main stream velocity. The isokinetic probe insertion depth must position the probe inlet at the pipe centreline for velocities above 2 m/s. Sample lines must be heated to at least 5°C above the pressure dew point at all points between the sample port and the measurement instrument to prevent condensation that would distort oil and moisture measurements. For oil content measurements per ISO 8573-2, sample line temperature must exceed the pressure dew point by at least 10°C.
±5% of main airstream velocity at probe inlet (ISO 8573-4)
Isokinetic Tolerance
Pipe centreline — velocities above 2 m/s
Probe Position
≥5°C above pressure dew point throughout sample line
Sample Line Temperature
≥10°C above PDP — prevents condensation distortion of oil content
Oil Sample Line
02 /
Solid Particle Measurement — ISO 8573-4
ISO 8573-4:2019 defines two methods for measuring solid particle content: gravimetric (mass-based, Method A) and particle counting (size-based, Method B). Method A collects particles on a pre-tared membrane filter with a known pore size (typically 0.8 µm); the collected mass divided by the sampled air volume gives contamination in mg/m³. This method measures the total mass of solid particles above the membrane cut-off pore size. Method B uses an Automatic Particle Counter (APC) calibrated per ISO 11171 to count and size particles at standard particle size ranges. Method B is used when the ISO 8573-1 purity class requires verification against particle count limits (classes 1–4 for solid particles). Sampling volume must be sufficient for reliable measurement: ISO 8573-4 recommends a minimum 1,000 L (at standard conditions 20°C, 100 kPa absolute) for gravimetric measurement when concentrations are near the detection limit of the membrane filter.
Membrane filter, pre-tared; result in mg/m³ — total mass above filter pore size
Method A (Gravimetric)
APC per ISO 11171; counts at defined size ranges — validates numerical limits
Method B (Particle Count)
1,000 L at standard conditions (20°C, 100 kPa abs)
Minimum Sample Volume
ISO 8573-4:2019 — supersedes 2001 edition
Standard Reference
03 /
Oil Aerosol Measurement — ISO 8573-2
ISO 8573-2:2018 defines the method for measuring oil aerosol content in compressed air. The method collects oil aerosol on a glass wool or borosilicate glass microfibre coalescing filter; the collected oil mass is determined by solvent extraction followed by gravimetric or GC-FID analysis. The minimum detectable concentration using gravimetric analysis is approximately 0.03 mg/m³; for ISO 8573-1 Class 1 verification (≤0.01 mg/m³), GC-FID analysis of the extract provides the sensitivity required. Minimum sample volume for Class 1 measurement: ≥10,000 L. The sampled air is directed through a heated sample line into the coalescing collector; the collector temperature must be maintained above the system pressure dew point. The measurement result is expressed as mg/m³ at reference conditions (20°C, 100 kPa absolute, 0% relative humidity). Combined oil content (aerosol plus vapour) is measured by adding the ISO 8573-2 aerosol result to the ISO 8573-5 vapour result.
ISO 8573-2:2018 — oil aerosol coalescing filter + gravimetric/GC-FID
Standard
Gravimetric: ~0.03 mg/m³; GC-FID: sufficient for ≤0.01 mg/m³
Class 1 Detection Limit
≥10,000 L at reference conditions
Sample Volume (Class 1)
20°C, 100 kPa abs, 0% RH — per ISO 8573-1 definition
Result Reference Conditions
04 /
Oil Vapour Measurement — ISO 8573-5
ISO 8573-5:2001 defines the method for measuring oil vapour and organic solvent content in compressed air. Vapour-phase oil compounds (typically C₁₁ to C₂₅ hydrocarbons from compressor lubricants) are collected by passing the air sample through a tube packed with activated charcoal (Tenax or activated charcoal adsorbent). After sampling, the adsorbent is thermally desorbed or solvent-extracted, and the eluate is analysed by Gas Chromatography with Flame Ionisation Detection (GC-FID). Results are expressed as total hydrocarbon concentration in mg/m³ at reference conditions. For ISO 8573-1 Class 1 and 2 verification, the combined total oil (aerosol + vapour) must not exceed 0.01 mg/m³ (Class 1) or 0.1 mg/m³ (Class 2) as a single combined result; ISO 8573-5 vapour and ISO 8573-2 aerosol results are summed. The adsorbent tube sampling flow rate and sample volume must be validated for the specific hydrocarbon compounds expected from the compressor oil formulation in use.
ISO 8573-5:2001 — activated charcoal adsorption tube + GC-FID
Standard
≤0.01 mg/m³ total (aerosol + vapour per ISO 8573-2 + ISO 8573-5)
Combined Oil Class 1
≤0.1 mg/m³ total (aerosol + vapour)
Combined Oil Class 2
C₁₁–C₂₅ typical compressor oil spectrum — adsorbent validated for compound class
Hydrocarbon Range
05 /
Viable Microbiological Contamination — ISO 8573-7
ISO 8573-7:2003 defines the method for measuring viable microbiological contamination (bacteria and mould) in compressed air. Two sampling methods are specified: an Anderson Cascade Impactor (for size-resolved viable aerosol characterisation) and an AGI-30 all-glass impinger (for total viable count). For routine ISO 8573-1 class verification, the impinger method is most commonly used. The AGI-30 impinger collects air samples into a sterile liquid collection medium at 12.5 L/min; minimum sample volume is 1,000 L (approximately 80 minutes at 12.5 L/min). After sampling, the collection medium is diluted in serial dilutions and plated on blood agar growth medium; plates are incubated at 37°C ±1°C for 48 hours before colony counting. Results are expressed in Colony Forming Units per cubic metre (CFU/m³). ISO 8573-7 microbiological purity classes: W1 ≤0.1 CFU/m³, W2 ≤1 CFU/m³, W3 ≤10 CFU/m³. Microbiological sampling requires aseptic technique throughout; sample equipment is autoclaved before use and collection liquid is sterile.
ISO 8573-7:2003 — impinger collection + agar plate culture (CFU/m³)
Standard
≤0.1 CFU/m³ — pharmaceutical and food-grade air quality
Class W1
≤1 CFU/m³ — medical and sensitive process air
Class W2
≥1,000 L at 12.5 L/min (≈80 min sampling) — minimum for reliable count
Sample Volume
ENGINEERING DIAGRAMS
ENGINEERING REFERENCES
ISO 8573-1:2010, Compressed Air — Part 1: Contaminants and Purity Classes
Purity class specification standard defining maximum allowable concentrations of particles, water, and oil in compressed air.
ISO 8573-2:2018, Compressed Air — Part 2: Test Methods for Aerosol Oil and Particle Content
Point-of-use oil and particle measurement test methods including impinger sampling, activated carbon adsorption, and particle counting procedures.
ISO 8573-3:2010, Compressed Air — Part 3: Test Methods for Measurement of Humidity
Dewpoint and humidity measurement methods for verifying ISO 8573-1 water class compliance at point of use.
EU GMP Annex 1:2022, Manufacture of Sterile Medicinal Products (Compressed Air Section)
EU pharmaceutical manufacturing guideline requiring compressed air quality management plans, continuous monitoring, and formal documentation for Grade A/B pharmaceutical air applications.
FREQUENTLY ASKED QUESTIONS
RELATED ENGINEERING TOPICS
CITE THIS PAGE
ELIMFILTERS. (2026). Compressed Air Quality Verification — ISO 8573-4, -5, and -7: Compressed Air Quality Verification — ISO 8573-4, -5, and -7. ELIMFILTERS Engineering Knowledge Platform. https://elimfilters.com/knowledge-center/engineering/compressed-air-quality-verification