Engineering · 10 min
Compressed Air Purity
ISO 8573-1 Quality Classes, Contamination Measurement, and System Design
Compressed air contains particles, liquid water, water vapour, and oil aerosols and vapour introduced during compression and distribution. ISO 8573-1 defines quality classes for each contaminant type, providing a standardised specification language for compressed air systems. Achieving the correct purity class requires understanding contamination sources, measurement methods, and the filtration and drying technologies that control each contaminant type.
≤0.01 mg/m³
ISO 8573-1 Class 1 oil
≤ −70°C
Class 1 pressure dew point
Zero permitted
Class 1 particles (≥5 µm)
+2°C to +10°C
Refrigerant dryer PDP
01 /
ISO 8573-1 Quality Classes
ISO 8573-1 defines quality classes 0 through 6 for three contamination categories independently: solid particles (by count and size), moisture (by pressure dew point, °C), and total oil (aerosol plus vapour, mg/m³). Class 0 is specified by the user and is more stringent than Class 1 — it is not a defined numerical value in the standard. Class 1 is the highest standard-defined purity; Class 6 is the least stringent. A compressed air quality specification written as ISO 8573-1:2010 [2:4:2] means Class 2 for particles, Class 4 for moisture, and Class 2 for oil. Each class must be specified independently because different treatment stages address each contamination type — a highly efficient particulate filter provides no benefit for dew point, and a refrigerant dryer reduces moisture but does not remove oil aerosols.
User-defined, more stringent than Class 1
Class 0
≤0.01 mg/m³ (total oil)
Class 1 oil content
ISO 8573-1 [2:4:2] = particles:moisture:oil
Notation example
02 /
Particle Quality Classes
ISO 8573-1 particle classes are defined by the maximum number of particles per cubic metre at specified size ranges. Class 1 (particles): ≤20,000 particles/m³ at 0.1–0.5 µm; ≤400 at 0.5–1 µm; ≤10 at 1–5 µm; no particles ≥5 µm. Class 2: same ≤400,000 at 0.1–0.5 µm; ≤6,000 at 0.5–1 µm; ≤100 at 1–5 µm; no particles ≥5 µm. Class 3: no limit for <1 µm; ≤1,000 at 1–5 µm; no particles ≥5 µm. Classes 4 and 5 allow particles ≥5 µm in increasing quantities. Class 6 allows particle mass concentration up to 10 mg/m³ without size distribution limits. For most industrial pneumatic actuator and tool applications, particle Class 3 or 4 is adequate. Instrument air (analytical instruments, pressure transmitters, control valves) requires particle Class 1 or 2 to prevent instrument blockage and measurement error.
Zero permitted
Class 1 (≥5 µm particles)
≤1,000 particles/m³
Class 3 (1–5 µm)
Class 1 or 2
Instrument air class
03 /
Moisture Classes and Pressure Dew Point
Moisture in compressed air is expressed as pressure dew point (PDP) — the temperature at which condensation occurs at the system operating pressure. A PDP of −40°C at 7 bar means the air contains so little moisture that condensation will not form until the temperature drops to −40°C, even at 7 bar line pressure. ISO 8573-1 moisture classes: Class 1: PDP ≤ −70°C; Class 2: ≤ −40°C; Class 3: ≤ −20°C; Class 4: ≤ +3°C; Class 5: ≤ +7°C; Class 6: ≤ +10°C. Refrigerant dryers achieve PDP between +2°C and +10°C (Classes 4–6). Adsorption (desiccant) dryers achieve PDP between −20°C and −70°C (Classes 1–3). The conversion from PDP to water content in g/m³ at a reference condition allows calculation of absolute humidity: at −40°C PDP and 7 bar, the water content is approximately 0.003 g/m³ — compared to 1.4 g/m³ at +7°C PDP.
≤ −70°C
Class 1 PDP
+2°C to +10°C (Classes 4–6)
Refrigerant dryer PDP range
−20°C to −70°C (Classes 1–3)
Desiccant dryer PDP range
04 /
Oil Content Classes and Measurement
Oil in compressed air originates from lubricant carryover from oil-lubricated compressors (in the form of aerosol and vapour) and from atmospheric hydrocarbon vapours that concentrate during compression. ISO 8573-1 total oil classes (aerosol + liquid + vapour, mg/m³): Class 1: ≤0.01 mg/m³; Class 2: ≤0.1 mg/m³; Class 3: ≤1 mg/m³; Class 4: ≤5 mg/m³. Oil-free compressors (Class 0 equipment per ISO 8573-7) still require oil removal filtration for Class 1 or Class 2 applications because atmospheric hydrocarbon vapours drawn into the compressor inlet concentrate to measurable levels at the outlet. ISO 8573-2 defines the test method for aerosol oil measurement: impaction on a membrane filter with gravimetric analysis. ISO 8573-5 defines the vapour oil measurement method using activated carbon adsorption tubes. Total oil = aerosol (ISO 8573-2 method) + oil vapour (ISO 8573-5 method). Coalescing filters remove aerosol oil to Class 1; activated carbon adsorption removes vapour oil — both stages are required to achieve verified Class 1 compliance.
≤0.01 mg/m³
Class 1 oil content
Coalescing filter (ISO 8573-2)
Aerosol removal
Activated carbon (ISO 8573-5)
Vapour removal
05 /
Contamination Sources During Compression
Ambient air drawn into a compressor contains atmospheric particulates (0.1–10 µm pollen, dust, soot), water vapour (humidity), and trace hydrocarbon vapours. During compression, all constituents are concentrated in proportion to the compression ratio: at 7 bar gauge (8 bar absolute), volume reduces to 1/8 and contaminant mass per volume increases 8-fold. Compressed air after-cooling (required to prevent thermal damage to downstream equipment) causes the compressed, humid air to cool rapidly — water vapour condenses into liquid water and aerosol droplets. The after-cooler and condensate separator remove the largest water fraction (typically 70–90% of the inlet water vapour at tropical ambient conditions) but leave substantial residual moisture requiring drying. Oil-lubricated compressors inject oil into the compression stage for cooling and sealing — oil carryover from the separator and coalescing filter in the compressor unit outlet is typically 1–5 mg/m³ before point-of-use treatment.
06 /
Filtration System Design for Quality Classes
A compressed air treatment train is designed in stages, each addressing one contamination category at the conditions produced by the upstream stage. After-cooler + condensate separator: removes bulk liquid water and coarse aerosols (Stages 4–5 moisture, particle Class 4–5). General purpose coalescing filter (1 µm coalescence element): removes liquid aerosol oil to 0.1 mg/m³ and particles ≥1 µm (particle Class 2–3, oil Class 2–3). High-efficiency coalescing filter (0.01 µm coalescence element): reduces oil aerosol to 0.01 mg/m³ (oil Class 1). Activated carbon adsorber: removes oil vapour below 0.003 mg/m³ (oil Class 1). Desiccant dryer: achieves PDP ≤ −40°C (moisture Class 2). Final particulate filter (PTFE membrane or borosilicate fiber): ensures sterile or classified air downstream of all treatment stages (particle Class 1 or 2). DRYCORE™ filtration systems provide integrated particulate and coalescing elements for compressed air applications, reducing system footprint and pressure drop budget.
Oil Class 2 (0.1 mg/m³)
Coalescing filter achieves
Oil Class 1 (0.01 mg/m³)
High-efficiency coalescer
Vapour removal to full Class 1 compliance
Activated carbon adds
ENGINEERING DIAGRAMS
COMMON ENGINEERING MISTAKES
Specifying ISO 8573-1 Class 0 air quality when the application only requires Class 1 or Class 2. Class 0 requires an inherently oil-free compressor — applying it to systems served by lubricated compressors forces complete system replacement rather than downstream treatment.
Measuring compressed air quality only at the dryer or filter outlet. Compressed air picks up contamination from distribution pipework — point-of-use measurements at the application connection are required to verify actual delivered quality.
Confusing pressure dew point (PDP) with atmospheric dew point. ISO 8573-1 water classes specify PDP at line pressure — a PDP of −40°C at 7 bar corresponds to approximately −27°C atmospheric dew point; the two figures are not interchangeable in specifications.
ENGINEERING REFERENCES
ISO 8573-1:2010, Compressed Air — Part 1: Contaminants and Purity Classes
Primary standard defining contamination classes for particles, water (dewpoint), and oil in compressed air systems.
ISO 8573-2:2018, Compressed Air — Part 2: Test Methods for Aerosol Oil and Particle Content
Test methods for measuring oil aerosol and particle concentration in compressed air at point-of-use sampling.
ISO 8573-3:2010, Compressed Air — Part 3: Test Methods for Measurement of Humidity
Test methods for pressure dewpoint measurement used to verify ISO 8573-1 water class compliance.
ISO 8573-7:2003, Compressed Air — Part 7: Test Methods for Viable Microbiological Contaminants
Microbiological testing standard for breathing air and pharmaceutical-grade compressed air applications.
FREQUENTLY ASKED QUESTIONS
RELATED ENGINEERING TOPICS
Air Intake System Design
Engineering
Airflow Engineering
Engineering
Compressed Air Dryer Selection: Refrigeration, Desiccant, and Membrane Drying Technologies vs ISO 8573-1 Dew Point Classes
engineering
Compressed Air Quality Verification — ISO 8573-4, -5, and -7
Engineering
Crankcase Ventilation and Blow-By Gas Filtration
Engineering
Filter Media Science
Engineering
CITE THIS PAGE
ELIMFILTERS. (2026). Compressed Air Purity: Compressed Air Purity. ELIMFILTERS Engineering Knowledge Platform. https://elimfilters.com/knowledge-center/engineering/compressed-air-purity